Run a dilution refrigerator.
Use the simulated controller to follow helium from storage to circulation, and see where millikelvin cooling comes from.
Controller
Ready to evacuate the insulating vacuum can.
Circulation controls
LockedRelative training controls. Available in circulation.
Where the helium goes
Ready at room temperature
The dilution interface
The coldest stage is where ³He crossing into the dilute liquid absorbs heat.
Simulated temperature history
Model scope and sources
The topology is a functional LD-family teaching schematic. The controller actions are training presets. Valve names, timing, and interlocks are illustrative. Real equipment follows its own controller, configuration, and lab procedures.
Temperatures come from a reduced thermal model and run on accelerated time. The model uses relative mixture inventory and flow. It does not resolve pressures, pump curves, composition evolution, detailed precool cycles, or hardware interlocks. The pulse tube and the dilution mixture are separate closed circuits. The vacuum circuit removes air from the insulating can.